All fluid flow here obeys Poiseuille's Law for laminar flow through a tube:
Q = (π × r⁴ × ΔP) / (8 × η × L)Where Q = flow rate, r = radius, ΔP = pressure difference, η = fluid viscosity (saline ≈ 0.001 Pa·s), L = tube length. The r⁴ term is critical — doubling the radius increases flow 16×. A tiny change in gap has an enormous effect.
Saline pumped through the scope's working channel behaves as flow through a tube or annular space (if a laser is present). The working channel diameter is user-configurable (default 3.6 Fr); most instruments have effective diameters 3.0–3.6 Fr. If a laser occupies part of the channel, fluid flows in the annular gap around it. The driving pressure is pump pressure minus whatever IRP already exists in the kidney:
Q_in = K_in × (P_pump − IRP) K_in = (π × r_wc⁴) / (8 × η × L_scope) [or annular formula if laser present]Fluid escapes through the ring-shaped gap between the scope outer wall and the sheath inner wall. This is not a simple circle — it's an annulus, so the flow factor uses the exact annular Poiseuille formula:
Q_out = K_out × (IRP − (−P_suction)) K_out = (π × F_annular) / (8 × η × L_sheath) F_annular = (r_o⁴ − r_i⁴) − (r_o² − r_i²)² / ln(r_o/r_i)Outflow is driven by the pressure difference: kidney pressure (IRP) minus the negative pressure from the suction pump (−P_suction). This is equivalent to IRP plus the suction magnitude. The annular factor F is orders of magnitude smaller than r⁴ for a solid tube of the same outer radius — this is why the gap is the bottleneck, not the sheath size.
At equilibrium inflow must equal outflow. Setting Q_in = Q_out and solving for the suction pressure that achieves your IRP target:
K_in × (P_pump − IRP) = K_out × (IRP − (−P_suction)) P_suction = K_in × (P_pump − IRP) / K_out − IRPThe calculator solves this in real time. The result is then multiplied by the clinical safety factor (default 1.5×) to account for non-ideal real-world conditions (pulsatility, scope deflection, debris). If the theoretical P_suction (before safety factor) comes out negative, passive drainage alone is sufficient — no Neptune needed.
Even a 1Fr increase in sheath ID widens the gap by ~0.17mm on each side. Because of the r⁴ relationship inside F_annular, this small change can double or triple outflow capacity, dramatically cutting the Neptune pressure required. This is the physical basis for using 12–14Fr sheaths with 9Fr scopes.
This model assumes laminar, steady-state flow of Newtonian fluid (saline). Key limitations:
Working channel diameter — Most flexible ureteroscopes have effective hydraulic diameters slightly smaller than nominal spec. Default is 3.6 Fr; try reducing to 3.0–3.2 Fr for conservative estimates.
Scope position — Scope deflection, eccentric positioning, and debris reduce real-world outflow. The annular flow factor may be 20–40% higher in practice.
Pulsatile flow & transients — This model solves steady-state equilibrium. Real pump surges and outflow lag during irrigation bursts mean peak IRP spikes above the equilibrium prediction.
For these reasons, the clinical safety factor (default 1.5×) multiplies the calculated suction to provide margin. Verify margins empirically under real operating conditions.